Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US Softw
FDA device recall Z-2518-2026 · posted 2026-06-18 · Open, Classified
Recall details
- Recalling firm
- Thoratec LLC
- Reason for recall
- Due to 11 volt Backup Battery failures.
- Action taken
- On 05/13/2026, the firm sent email an "URGENT MEDICAL DEVICE RECALL" Letter informing customers that As of April 27, 2026, Abbott received twenty-four (24) complaints where an advisory warning, Replace Backup Battery, was triggered once certain 11V Backup Batteries were connected to the System Controller. This warning is displayed as a yellow banner on the HeartMate Touch Monitor and System Monitor. This advisory alert is only visible when using HeartMate Touch or the System Monitor in the clinic, it is not visible to the patient. Customers are instructed to: 1. Read this notification thoroughly, complete and sign the form included with this letter, acknowledging that you received and understand this information, and that you will communicate this information to those relevant in your facility. Once complete, return the acknowledgement form to Abbott. 2. Use the online lookup tool (https://www.cardiovascular.abbott/us/en/hcp/product-advisories/heartmatebatteries-2026.html) to confirm if a 11V Lithium-Ion battery at your site is an impacted one. Figure 3 shows an instruction on how to locate the serial numbers of the impacted batteries. 3. Once the impacted batteries are identified, do not use them. 4. Beginning May 15, 2026, a separate email will be sent to your site confirming shipment of replacement batteries. This email will include return shipping labels and a device retrieval form, along with detailed instructions for returning the affected batteries. For Questions - contact your Abbott representative or Abbott Technical Support at 1-800-456-1477 (Business hours: 8 AM EST to 7 PM EST).
- Root cause
- Under Investigation by firm
- Status
- Open, Classified
- Product code
- DSQ
- Quantity affected
- 4
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA
- Date initiated
- 2026-05-13
- Date posted
- 2026-06-18
- Location
- Pleasanton, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.