Brand Name: EMPOWR 3D KNEE TIBIAL INSERT - E-PLUS Product Name: EMPOWR 3D KNEETM INS, 11L 10MM, VE Model/Catalog Number:
FDA device recall Z-2125-2025 · posted 2025-07-17 · Open, Classified
Recall details
- Recalling firm
- Encore Medical, LP
- Reason for recall
- Due to packaging discrepancy with the incorrect tibial inserts/implants being packaged. The use of this product may result in the incorrect product being selected. This could result in a revision surgery if used.
- Action taken
- On 06/02/2025, the firm emailed an "Urgent: Medical Device Recall" letter to customers. The letter described the product, problem and actions to be taken. Customers are instructed to: 1.Review their stock for the products and lot numbers for the items to be returned. 2.This notice needs to be passed on to all those within their organization or to any organization where the product has been transferred. 3.Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 or through their regional email address, to place a replacement request. They will be provided with an RMA number and shipping label to return the affected product. 4.Return all devices (quantities listed above) using the RMA number and shipping label provided by Customer Service. Any questions, contact Recall Coordinator at productsafety@enovis.com or call/text 682-299-9133.
- Root cause
- Packaging change control
- Status
- Open, Classified
- Product code
- JWH
- Quantity affected
- 20
- Distribution
- U.S Distribution to states of: AZ, IA, ID, IL, IN, LA, MI, MO, MS, NE, NY, OK, OR, PR, SD, UT, VA, and WA
- Date initiated
- 2025-06-02
- Date posted
- 2025-07-17
- Location
- Austin, TX