Atlas FDA

Brand Name: Daytona Small Stature Product Name: Ti Locking Cap 4.5mm, Small Stature, T25 Model/Catalog Number: 79-0001 S

FDA device recall Z-0752-2025 · posted 2024-12-23 · Open, Classified

Recall details

Recalling firm
SEASPINE ORTHOPEDICS CORPORATION
Reason for recall
Due to manufacturing defects observed on Locking Caps that affect the ability for components to mate.
Action taken
Update 12/18/2024 - The firm sent an updated Customer Notification Letter on 12/12/2024 to customers that have not responded to the initial notification on 10/29/2024. On 10/29/2024, firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter to customer informing them that the firm has identified a nonconformance with the threads of the locking cap components that may be incongruous with optimal manufacturing specifications. The nonconformance can result in a locking cap that does not mate appropriately with the screw tulip. Customers are instructed to: -Review their inventory and return the affected part(s) using the provided label to return the inventory in their possession. Any questions contact Dalton Daggs at Dalton.daggs@seaspine.com or Danielle Nelson-Archer at Danielle.nelson-archer@seaspine.com.
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
NKB
Quantity affected
191 devices
Distribution
U.S. Nationwide distribution in the states of CA, CO, FL, GA, PA, TX, and VA.
Date initiated
2024-10-29
Date posted
2024-12-23
Location
Carlsbad, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
NewPort Spinal System; Daytona Small Stature Spinal System; Malibu Spinal System; Mariner Pedicle Screw Systems: Mariner Outrigger Revision System; Mariner MIS Pedicle Screw System; Mariner Deformity System; Mariner RDX System (K232566)SeaSpine Orthopedics Corporation2024-04-19