Brand Name: da Vinci S, Si Mega Needle Driver Instrument Product Name: da Vinci S, Si Mega Needle Driver Instrument Mode
FDA device recall Z-1698-2026 · posted 2026-03-30 · Open, Classified
Recall details
- Recalling firm
- Intuitive Surgical, Inc.
- Reason for recall
- Due to increased complaints for broken/frayed grip cables for reusable instruments.
- Action taken
- On 12/29/2025, the firm sent via email an "Urgent: Medical Device Recall" Letter informing customers that intuitive has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci S and Si reusable instruments. Customers are instruct to: 1. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 2. Identify and quarantine any affected product(s). Affected product(s) listed above may be returned by sending an email with quantities and lot numbers to your regional customer service. 3. Ensure to include the FSCA number ISIFA2025-15-R in the return notes. 4. Credit will be provided based on the number of remaining lives. 5. Shared or further distributed these products with other sites, make sure appropriate staff at the site receive and understand this notification so they locate and return their affected product. For questions or concern, contact Clinical Sales Representative or contact Intuitive Customer Service at the numbers listed below: " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support.korea@intusurg.com " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or csjapan@intusurg.com " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)
- Root cause
- Process control
- Status
- Open, Classified
- Product code
- NAY
- Quantity affected
- 7819
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MO, MS, NC, NJ, NY, OH, PA, TX, VA and WI. The countries of Argentina, Brazil, Canada, Chile, China, C
- Date initiated
- 2025-12-09
- Date posted
- 2026-03-30
- Location
- Sunnyvale, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM, MODEL IS2000 (K050369) | Intuitive Surgical, Inc. | 2005-04-29 |