FDA device recall Z-2906-2026 · posted 2026-08-06 · Open, Classified

Recall details

Recalling firm
TISSIUM SA
Reason for recall
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Action taken
On June 25, 2026, the firm began notifying consignees via direct visit by medical representatives to deliver the Customer Recall Notification Letter in person. Representatives removed affected units from customer shelves, collected completed response forms, and provided conforming replacement product on the same day.
Root cause
Incorrect or no expiration date
Status
Open, Classified
Product code
SFD
Distribution
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Date initiated
2026-06-15
Date posted
2026-08-06
Location
France