Atlas FDA

Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model

FDA device recall Z-0304-2026 · posted 2025-10-24 · Open, Classified

Recall details

Recalling firm
Alcon Research LLC
Reason for recall
Due to incomplete seals in the pouch which provide the sterile barrier.
Action taken
On 08/05/2025, the firm sent via FedEx Overnight an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that for the Alcon Standalone Vitrectomy Consumables, there is potential for some pouches within impacted lots to have an incomplete seal. Customers are instructed to: locate and dispose of any affected lots of Alcon Standalone Vitrectomy Consumables remaining in your inventory. To comply with this Medical Device Recall and request the replacement of any unused product, please take the following steps: 1. Review inventory to determine if there are any unused affected product within their facility. See table on page 1 for affected Standalone Vitrectomy Consumable lots shipped to your location. 2. Segregate and dispose of any unused affected product from their inventory. 3. Call Alcon Customer Service to arrange for replacement of the affected inventory of Alcon Standalone Vitrectomy Consumables. 4. Respond to Alcon indicating understanding of these instructions even if they have zero (0) units remaining in inventory by completing the online response form at https://qrco.de/Vit-Seal or by completing and returning the attached Response Form and returning to Alcon via email: Market.Actions@Alcon.com or fax: 817-302-4337. 5. Forward this notification to all departments within their organization who may be in possession of this affected product; and any other organization to which this product may have been transferred. For questions contact, Alcon Customer Service at 1-800-862-5266 or an Alcon Sales Representative.
Root cause
Process control
Status
Open, Classified
Product code
MLZ
Quantity affected
577
Distribution
Worldwide distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC,SD,T
Date initiated
2025-08-05
Date posted
2025-10-24
Location
Fort Worth, TX

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Alcon 27GA Chandelier (8065000252); Alcon 27+ DS Wide Angle Endoilluminator (8065000253); Alcon 27+ DS Endoilluminator (8065000256); (K242184)Alcon Laboratories, Inc.2024-12-04
UNITY VCS (8065000296); UNITY CS (8065000297) (K233876)Alcon Laboratories, Inc.2024-06-21
CONSTELLATION VISION SYSTEM (K141065)Alcon Research, Ltd.2014-08-22