BP7561 OPTI CCA-TS B-Lac Cassette (25 per box) Cassettes are consumables used in the OPTI CCA-TS Analyzer intended to be

FDA device recall Z-2440-2012 · posted 2012-09-21 · Terminated

Recall details

Recalling firm
OPTI Medical Systems, Inc
Reason for recall
The values reported for PCO2 by BP7561 B-Lac cassettes are biased at high concentrations when used at altitudes above 2800 feet (853 meters).
Action taken
OPTI Medical sent a Urgent Field Correction Notice dated July 3, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to apply correlation factors to correct the PCO2 measurement reported by the device. They were instructed to determine the elevation (altitude above sea level) for their location and to use the tables provided in the recall notification to determine the correct correlation factor. A website was provided if the consignee did not know their elevation. Steps were provided to enter the correlation factor into their OPTI CCA TS analyzer. Please contact OPTI Medical Technical Support at 1-770-510-4444 or 1-800-490-6784.
Root cause
Process design
Status
Terminated
Product code
CHL
Quantity affected
895 boxes (25 cassettes per box)
Distribution
Worldwide Distribution -- US (nationwide) including the states of AZ, ID, TX and UT., and the countries of Argentina, Bolivia, China, Ecuador, France, Germany Greece, India, Indonesia, Iran, Ireland,
Date initiated
2012-07-03
Date posted
2012-09-21
Date terminated
2013-04-11
Location
Roswell, GA