Boston Scientific SMARTFREEZE CRYO CONSOLE: Console: Material Number M004CRBS4000; CONSOLE REFURB: Material Number M004C

FDA device recall Z-0303-2025 · posted 2024-11-12 · Open, Classified

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
Action taken
Boston Scientific disseminated a letter to Physicians/Healthcare Professionals on 10/10/2024 via courier. The letter notified its consignees of updates to the instructions for use (IFU) for the POLARx and POLARx FIT Cryoablation Catheter related to the risk of atrioesophageal (AE) fistula. The firm requested that the information be shared with clinicians who use the use the Boston Scientific POLARx Cryoablation System, including the POLARx Catheter, the POLARx FIT Catheter and the SMARTFREEZE Console, and to share the communication with any other organization to which these devices may have been transferred. Users were directed to immediately post the information on or near the product to ensure this information is easily accessible to all users of the device.
Root cause
Device Design
Status
Open, Classified
Product code
OAE
Quantity affected
1035 units
Distribution
Worldwide
Date initiated
2024-10-10
Date posted
2024-11-12
Location
Saint Paul, MN