Atlas FDA

Boston Scientific LeVeen SuperSlim Needle Electrode, 3.0 cm/15 cm, sterile, Boston Scientific, Spencer, IN; REF 26-228,

FDA device recall Z-1915-2008 · posted 2008-09-11 · Terminated

Recall details

Recalling firm
Boston Scientific Corp
Reason for recall
The cannula may become detached from its correct orientation inside the handle and may prevent retraction of the tines.
Action taken
Consignees were notified by recall letter on 4/30/08, sent fexex priority mail, to cease use of the products and to return them to the firm. For assistance, contact Boston Scientific at 1-508-683-4186.
Root cause
Device Design
Status
Terminated
Product code
GEI
Quantity affected
9,843
Distribution
Nationwide, Argentina, Belgium, Canada, Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hong Kong, Hungary, Iraq, Ireland, Italy, Jordan, Korea, Lebanon, Malaysia, Netherlands, New Zea
Date initiated
2008-04-30
Date posted
2008-09-11
Date terminated
2009-08-04
Location
Spencer, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
LEVEEN NEEDLE ELECTRODE (K982556)Radiotherapeutics Corp.2000-04-06