Boston Scientific IQ Guide Marker Wire. Catalog No. 38951-01J0, 185 cm, J-Tip, Single Pack. Catalog No. 38951-01J2, 185
FDA device recall Z-1218-06 · posted 2006-07-12 · Terminated
Recall details
- Recalling firm
- Boston Scientific
- Reason for recall
- Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.
- Action taken
- On May 30, 2006, Boston Scientific notified customers of the IQ Guide Wire manufactured from February 3, 2006, through March 15, 2006. BSC requested a reply whether or not they still have product in their institution. Replacements will be issued for returned product. UPDATE: A second recall notification was sent to affected customers 07/07/06. This second notification included Inner Pouch Number (UPN) and Inner Pouch Lot/Batch #.
- Root cause
- Other
- Status
- Terminated
- Product code
- DQX
- Quantity affected
- 1455 units total
- Distribution
- Nationwide (AR, AZ, FL, GA, IA, ID, IL, IN, KY, LA, MA, MI, MN, MO, NC, NH, NJ, NM, NV, OH, OK, PA, TN, TX, UT, VA, WI, WV) and Puerto Rico. Chile, Germany, Taiwan.
- Date initiated
- 2006-05-30
- Date posted
- 2006-07-12
- Date terminated
- 2007-06-28
- Location
- Maple Grove, MN