Boston Scientific Innova Over-the-Wire Self-Expanding Stent System, Catalog No. 39181-08023. 8 mm x 20 mm x 130 cm (sten

FDA device recall Z-2532-2011 · posted 2011-06-16 · Terminated

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Boston Scientific Corporation is conducting a recall of the Innova Self-Expanding Stent System because they have received 6 complaints involving no deployment/partial deployment of the Innova Stent. Potential health hazard events resulting from this type of failure include increased procedure time, vessel wall injury, and/or emergency surgery to remove the partially deployed stent. This recall d
Action taken
Boston Scientific sent an Urgent Medical Device Recall Removal - Immediate Action Required letter, dated May 13, 2011, to all affected customers via registered mail. The letter explained the issue, identified affected product, stated that further distribution and or use of the devices should cease immediately, and asked that devices in inventory to be segregated and immediately returned to Boston Scientific. An Account Reply Verification Tracking Form was also asked to be returned. Boston Scientific Sales Representative were assist in the return process by providing a Returned Goods Authorization (RGA) number. Customers were instructed to contact their local Sales Representative for any questions.
Root cause
Device Design
Status
Terminated
Product code
NIP
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Product removal in US is being managed through SuperNOVA Clinical Trial suspension in accordance with IDE #G100291. OUS distribution to i
Date initiated
2011-05-19
Date posted
2011-06-16
Date terminated
2014-02-04
Location
Maple Grove, MN