FDA device recall Z-1336-2015 · posted 2015-03-27 · Terminated

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Boston Scientific is initiating this Medical Device Field Correction of the Expel" APD and APDL Drainage Catheters. To date, Boston Scientific has received one complaint for device fragmentation after an Expel" APD catheter was implanted in the biliary system. The most serious foreseeable patient risk for this issue is additional intervention for fragment retrieval using minimally invasive meth
Action taken
Boston Scientific sent an "Urgent Medical Device Correction" letter dated February 25, 2015, to all affected customers. The letter described the problem and the product involved in the recall. The letter informed customers not to use the Expel APD and APDL Drainage catheters for bile drainage and to complete and return the Customer Acknowledgement Form. The letter also noted that the devices are not being retrieved and they are not required to return them to Boston Scientific. For further questions please call (763) 494-7971.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
FGE
Quantity affected
918
Distribution
Worldwide Distribution - US Distribution to the states of : CA, LA, FL, WI, NY, IL, MD and OH., and to the countries of: Belgium, Czech Republic, Finland, France, Germany, Great Britain, Italy, Spain,
Date initiated
2015-02-25
Date posted
2015-03-27
Date terminated
2016-08-12
Location
Maple Grove, MN