Boston Scientific brand Zero Tip Nitinol Stone Retrieval Basket, 12 mm, 90 cm, 1.9 Fr., sterile, latex free; Catalog no.
FDA device recall Z-0510-06 · posted 2006-02-08 · Terminated
Recall details
- Recalling firm
- Boston Scientific Corp
- Reason for recall
- The basket wire assembly may detach from the drive wire while in the patient.
- Action taken
- U.S. and foreign consignees were notified via letter dated 12/20/05 and requested to return stocks on hand. Distributors were requested to notify their customers.
- Root cause
- Other
- Status
- Terminated
- Product code
- FFL
- Quantity affected
- 2076
- Distribution
- Nationwide, Argentina, Australia, Belgium, Brazil, Canada, Chile, Denmark, Dominican Republic, Finland, France, Germany, Great Britain, Greece, Hong Kong, Ireland, Italy, Malaysia, Morocco, Netherland
- Date initiated
- 2005-12-20
- Date posted
- 2006-02-08
- Date terminated
- 2007-05-01
- Location
- Spencer, IN