Boston Scientific, Atlantis SR Pro 2, 40 MHz Coronary Imaging Catheter, 3.6F (1.18mm) x 135 cm, USE BY 2011-05, REF 3901
FDA device recall Z-2419-2011 · posted 2011-06-10 · Terminated
Recall details
- Recalling firm
- Boston Scientific Corporation
- Reason for recall
- Catheter tip detachment that is not associated with elongation of the catheter typically caused by excessive pulling forces. EXPANDED: Boston Scientific is initiating a recall of all iCross Coronary Imaging Catheters because they have received and confirmed 8 occurrences of catheter tip detachments due to embrittlement.
- Action taken
- Boston Scientific " Important Customer Notification" letters dated March 14, 2011 were hand delivered to customers by Boston Scientific sales representatives starting on March 28th, 2011. The sales representatives verbally reviewed information. A tracking log will be maintained document completion of the customer notifications. The letter described the product, problem and possible root cause as well as one corrective action, a change in sterilization method from Gamma to E-beam implementation. Boston Scientific sent an Urgent Medical Device Recall Removal - Immediate Action Required letter dated May 26, 2011, to all affected customers. Customers were instructed to Identify and discontinue use of any affected product within their inventory and to segregate it immediately and return it to Boston Scientific
- Root cause
- Process design
- Status
- Terminated
- Product code
- DQO
- Quantity affected
- 30,669
- Distribution
- Worldwide Distribution - USA (nationwide) and Puerto Rico and the countries of Japan, Trinidad, Tobago, and American Virgin Islands.
- Date initiated
- 2011-03-28
- Date posted
- 2011-06-10
- Date terminated
- 2012-07-30
- Location
- Maple Grove, MN
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| MODIFICATION TO ATLANTIS SR PRO 2 AND SR PRO CORONARY IMAGING CATHETERS, MODEL 39014 AND 38942 (K063312) | Boston Scientific Corp | 2006-11-30 |
| ATLANTIS SR PRO2 CORONARY IMAGING CATHETER, MODEL 39014 (K050577) | Boston Scientific Corp | 2005-03-30 |