Atlas FDA

Boston Scientific Advanix Pancreatic Stent, Straight Leading Barb, 10F x 18cm UPN: M00536730

FDA device recall Z-0560-2020 · posted 2019-11-25 · Terminated

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
Action taken
Boston Scientific issued Notifications to consignees via Federal Express Priority mail on October 21, 2019. The notification instructs the account to check their inventory remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form.
Root cause
Incorrect or no expiration date
Status
Terminated
Product code
FGE
Quantity affected
35 units
Distribution
Nationwide Foreign: Austria Australia Belgium Bulgaria Croatia Czech Rep Denmark Finland France Germany Great Britain Italy Japan Pakistan Poland Russian Fed. Slovakia South Africa Spain Sweden Switze
Date initiated
2019-10-21
Date posted
2019-11-25
Date terminated
2023-12-20
Location
Marlborough, MA