Boheringer Suction Regulator Model 7702L, Lot number 04D Boehinger Replacement Gauge, Model 9071, Lot number 200405.
FDA device recall Z-1231-04 · posted 2004-07-20 · Terminated
Recall details
- Recalling firm
- Boehringer Laboratories Inc
- Reason for recall
- The suction regulator incorporates a gauge for the verification of suction levels. The particular lot of gauges used in the production of the regulators exhibits a potential to develop a leak and cause the delivered suction to be inaccurate.
- Action taken
- All direct accounts were contacted by the recalling firm''s Customer Service Representative to determine the location of the affected product. The CSR arrange for replacement proudct.
- Root cause
- Other
- Status
- Terminated
- Product code
- GCX
- Quantity affected
- 56 units
- Distribution
- The recalled product was distributed to the following states: AL, FL, MA, MI, OR, PA, TN,
- Date initiated
- 2004-05-19
- Date posted
- 2004-07-20
- Date terminated
- 2004-09-14
- Location
- Norristown, PA