Atlas FDA

Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzer, ABL-700 series.

FDA device recall Z-1317-06 · posted 2006-08-02 · Terminated

Recall details

Recalling firm
Radiometer America Inc
Reason for recall
Software defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.
Action taken
The customers were apprised of this situation with a recall letter issued by the firm on 6/17/2005. In addition, new Operators Manuals were sent to the customers. In the letter, the customers were advised that a Service Engineer would be contacting them shortly in order to arrange for the installation of the software upgrade. The firm''s Service Engineers visited each customer location and installed software upgrades. Version 3.831 software was installed on the 700 Series Analyzers and Version 4.14 was installed on the 800 Series Analyzers. According to the firm, these software corrections were verified as having been completed by 3/10/2006.
Root cause
Other
Status
Terminated
Product code
CHL
Quantity affected
1635 devices
Distribution
Nationwide.
Date initiated
2005-06-17
Date posted
2006-08-02
Date terminated
2012-02-24
Location
Westlake, OH

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ABL-700 SERIES, MODEL ABL 7XX (VARIOUS MODELS) (K002290)Radiometer Medical A/S2000-09-13
ABL700 SERIES ANALYZER WITH AUTOCHECK MODULE (K992859)Radiometer America, Inc.2000-01-10
ABL 735 ANALYZER (K991417)Radiometer America, Inc.1999-11-24