Atlas FDA

Black Widow 6.6mm x 20mm Screw, Part Number TBS020; Black Widow 6.6mm x 25mm Screw, Part Number TBS025; Black Widow 6.6m

FDA device recall Z-0887-2017 · posted 2016-12-22 · Terminated

Recall details

Recalling firm
Amendia, Inc
Reason for recall
There is a problem with the mating feature between the plate and screw which could cause the plate opening not to interfere enough with the plate and head of the screw.
Action taken
Amendia sent an Urgent Medical Device Recall letter dated November 22, 2016, to all affected consignees.. The letter requested that they examine inventory and quarantine the product. The letter also requested that they contact Amendia to confirm receipt of the letter and to obtain a Return Material Authorization for any product on hand. Customers with questions should call 1-855-755-3329. For questions regarding this recall call 404-610-7215.
Root cause
Process change control
Status
Terminated
Product code
KWQ
Quantity affected
96 units
Distribution
Nationwide Distribution to IN, LA, and TX
Date initiated
2016-11-22
Date posted
2016-12-22
Date terminated
2017-07-31
Location
Marietta, GA