Bivona Uncuffed Pediatric Flex Tend Plus Extra Length Tracheostomy Tube Sizes 4.0 - 6.0 (With a Straight Neck Flange) Pr
FDA device recall Z-0665-2012 · posted 2012-01-19 · Terminated
Recall details
- Recalling firm
- Smiths Medical ASD, Inc.
- Reason for recall
- Difficulty disconnecting accessories from the Bivona Uncuffed Neonatal, Pediatric and Flextend Tracheostomy Tubes TR3 connector, which may result in a delay in initiating therapy or patient discomfort. If excessive force is used to detach the accessory, this may result in decannulation of the tube which could result in serious injury or death; especially if no replacement tracheostomy tube is imm
- Action taken
- Smith Medical sent an Urgent Field Safety Notice dated November 28, 2011, via U.S. Mail to all affected consignees. The notice identified the product, description of problem, and actions to be taken. This notice included a Customer Information Bulletin and Urgent Field Safety Notice Confirmation Form. Customers were instructed to 1) identify all affected unused product in inventory and segregate it to a quarantine location; 2) complete and return the attached Confirmation Form by fax to 219-989-7259 or email to bivona.tr3@smiths-medical.com; and 3) distributors were instructed to stop distributing and provide Smith Medical with a list of their customers, including contact details, so that Smith Medical can notify the customers of this notice. For questions contact Smiths Medical Customer Service at 800-258-5361, Option 1.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- BTO
- Quantity affected
- US distribution 153,033 devices ; Foreign 21,490 devices
- Distribution
- Worldwide Distribution - USA Nationwide: AZ, CA, CO, CT,DE, FL, GA, HI, IL, IN, KS, KY, LA, SC, TX, OR, NJ, MO, UT, MI, NY, NM, MA, MN, MS, NC, ND, NE, NH, NV, OH, OK, OR, RI, SD, TN, VA, VT, WA, WI,
- Date initiated
- 2011-11-28
- Date posted
- 2012-01-19
- Date terminated
- 2014-10-17
- Location
- Gary, IN
Monitor recalls programmatically