Atlas FDA

BirthNet OB Data Management System 91861

FDA device recall Z-0115-06 · posted 2005-11-01 · Terminated

Recall details

Recalling firm
Spacelabs Medical Incorporated
Reason for recall
Potential for incorrect weight to be displayed.
Action taken
On 10/13/04 the firm sent letters to customers advising of the error, and that software to correct the problem will be released the end of October. Service personnel will install new software.
Root cause
Other
Status
Terminated
Product code
HGM
Quantity affected
44 devices
Distribution
Distribution was to 44 medical facilities/hospitals nationwide.
Date initiated
2005-10-13
Date posted
2005-11-01
Date terminated
2011-01-17
Location
Issaquah, WA