Atlas FDA

BioRad BioPlex 2200 Software, for BioPlex 2200 Instrument, an automated immunodiagnostic analyzer, Bio-Rad Laboratories,

FDA device recall Z-0575-2008 · posted 2008-04-22 · Terminated

Recall details

Recalling firm
Bio-Rad Laboratories Inc
Reason for recall
Incorrect values: An error was found in the BioPlex 2200 Software, Version 2.0 that results in assignment values being utilized for the BioPlex assay Calibrator Set that may be slightly different from those printed on the Value Assignment Data Sheet. Some values may be identical between the lot assignment combinations with the ANA Screen with MDSS kit.
Action taken
Bio-Rad sent a Voluntary Field Correction notice dated November 1, 2007. Consignees were notified by phone call on Nov. 13, 2007, followed up by facsimile (domestic). Sub-recall to be by e-mail message for consignees outside the US. Distribution of the product ceased on Nov. 2, 2007. Bio-Rad will provide Calibrator Set CDs with only one lot of calibrator assignment values until a new version of software is available.
Root cause
Software design
Status
Terminated
Product code
NVI
Quantity affected
85 units of calibrator lot 86280 and 16 units of calibrator lot 86290 were distributed.
Distribution
Worldwide Distribution - USA, Canada, France, and Germany.
Date initiated
2007-11-13
Date posted
2008-04-22
Date terminated
2008-07-23
Location
Hercules, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
BIOPLEX 2200 ANA SCREEN WITH MEDICAL DECISION SUPPORT SOFTWARE FOR USE WITH BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM (K043341)Bio-Rad Laboratories, Inc.2005-10-27