Biomet Modular Primary Tibial Tray Implants; Modular Interlok Primary Tray (size 75 mm); Item No. 141215; UDI: (01) 0088

FDA device recall Z-2177-2018 · posted 2018-06-01 · Terminated

Recall details

Recalling firm
Zimmer Biomet, Inc.
Reason for recall
One lot of 75mm tibial tray is potentially etched and labeled as a 79 mm tibial tray. This may lead to delay in surgery, or immediate revision surgery to remove an incorrectly sized tibial component after it has been cemented.
Action taken
On May 18, 2018, the firm notified affected distributors and user facilities of the recall through Urgent Medical Device Recall letters sent via email and FedEx. The letters informed affected consignees of the issue and stated that the affected units were distributed in March 2018. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to CorporateQuality.PostMarket@zimmerbiomet.com within three (3) days. b. For each return, send a copy of Attachment 1 to CorporateQuality.PostMarket@zimmerbiomet.com. c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to CorporateQuality.PostMarket@zimmerbiomet.com. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected
Root cause
Under Investigation by firm
Status
Terminated
Product code
JWH
Quantity affected
16
Distribution
US Distribution to the states of : FL, GA, IN, MA, NM, UT, and WI. Foreign distribution to Canada and Mexico
Date initiated
2018-05-18
Date posted
2018-06-01
Date terminated
2020-03-31
Location
Warsaw, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Biomet Knee Joint Replacement Prostheses (K171054)Biomet, Inc.2017-07-28
Biomet Tibial trays (K142933)Biomet, Inc.2014-12-23
OFFSET TIBIAL TRAY (K010212)Biomet, Inc.2001-02-14
MCK MAXIMUM CONGRUENT KNEE) KNEE SYSTEM (K915132)Biomet, Inc.1992-02-11