Biomet 3i Narrow Right Angle Large Driver Tip (Hexed). Driver tips are attached to a surgical drilling unit to carry com
FDA device recall Z-0266-2016 · posted 2015-11-12 · Terminated
Recall details
- Recalling firm
- Biomet 3i, LLC
- Reason for recall
- The RASH3N, a hexed, abutment screw driver tip has been labeled and distributed as RASQ3N, a square, abutment screw driver tip. Additionally, the RASQ3N received the RASH3N's laser marking "RASH3N" during manufacturing.
- Action taken
- Beginning on or about 08/10/2015 customers will be notified in writing via FEDEX and instructed to to check their inventory for the identified product and to return any unused quantiles along with the recall response. Biomet 3i will promptly issue replacement product.
- Root cause
- Labeling mix-ups
- Status
- Terminated
- Product code
- NDP
- Quantity affected
- 96 pieces
- Distribution
- AK, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KY, LA, MA, MD, MI, MS, NC, NE, NJ, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, and District of Columbia Argentina, Australia, Belgium, Canada, Colombia, Greece, Ho
- Date initiated
- 2015-08-10
- Date posted
- 2015-11-12
- Date terminated
- 2016-10-12
- Location
- Palm Beach Gardens, FL