BioMerieux VIDAS System in combination with Quality Control VIDAS* (QCV*) Product Usage: the Quality Control VIDAS (QCV)
FDA device recall Z-2429-2018 · posted 2018-05-09 · Terminated
Recall details
- Recalling firm
- bioMerieux, Inc.
- Reason for recall
- Increase of events affecting performance regarding pump clogging and displacement of the color-coded SPR label (DOT). These performance issues could contribute to erroneous results for any assay performed on the VIDAS and mini VIDAS system due to a decrease of pipetted volume.
- Action taken
- Subsidiaries and distributors were initially notified of the recall on approximately 01/04/2018. U.S. customers were notified via letter on approximately 01/22/2018. On 03/28/2018 a follow-up notification was issued world-wide to customers to expand the scope. Instructions included to increase the frequency of the Quality Control VIDAS (QCV) testing to weekly rather than monthly to reduce the period of potential retrospective analyses needed, continue to conduct a visual inspection of the SPR after each run (including patient sample test and QCV test) to ensure that the dot on the SPR is still in place, continue to perform External QC testing in accordance with laboratory policy and in compliance with Governmental Regulations & Accreditation Requirements, which includes CLIA, ensure the notification letter is distributed to and reviewed by all appropriate personnel, store the notification letter with the bioMrieux VIDAS and /or miniVIDAS documentation, and complete and return the Acknowledgement Form in Attachment A by Fax to confirm receipt of the notice. For further questions, please call (314) 731-8694
- Root cause
- Device Design
- Status
- Terminated
- Product code
- JJE
- Quantity affected
- 1,866 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA,
- Date initiated
- 2018-01-04
- Date posted
- 2018-05-09
- Date terminated
- 2022-04-14
- Location
- Durham, NC
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| MINIVIDAS INSTRUMENT (K923579) | Biomerieux Vitek, Inc. | 1993-02-12 |