Biolox Delta Femoral Head - The BIOLOX delta Ceramic Femoral Head is used in conjunction with compatible acetabular and
FDA device recall Z-1134-2015 · posted 2015-02-20 · Terminated
Recall details
- Recalling firm
- Zimmer Gmbh
- Reason for recall
- Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present, a periprosthetic infection is possible, which could lead to revision.
- Action taken
- On 12/17/2014, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-800-348-2759.
- Root cause
- Package design/selection
- Status
- Terminated
- Product code
- LZO
- Quantity affected
- 10340
- Distribution
- Domestic: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MS, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY Internation
- Date initiated
- 2014-12-17
- Date posted
- 2015-02-20
- Date terminated
- 2015-09-23
- Location
- Winterthur, Switzerland
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| BIOLOX DELTA CERAMIC FEMORAL HEAD (K071535) | Zimmer, Inc. | 2007-11-19 |