Atlas FDA

Biogel Skinsense Polyisoprene Surgical Gloves, Non-Latex, Sterile, (Size 8) Made in Malaysia

FDA device recall Z-1219-2007 · posted 2007-09-08 · Terminated

Recall details

Recalling firm
Molnlycke Health Care, Inc
Reason for recall
Sterility compromised (package integrity): The primary packaging for a portion of the lot may contain gloves that were not sealed properly.
Action taken
Consignees were notified by registered mail on 07/23/2007. The notice contained a Recall Response Form to confirm receipt of the notice and to verify if the product is still in inventory. The notice instructed to consignee to inspect their inventory for Biogel Skinsense Polyisoprene Surgical Gove, lot #07B026, Size 8. If this lot and size are in their inventory, they were instructed to discontinue use, quarantine the product and return the product to Molnlycke Healthcare.
Root cause
Packaging process control
Status
Terminated
Quantity affected
6,000 pair
Distribution
Nationwide including states of AZ, CA, CT, KY, MD, MN, MO, TN, and WI
Date initiated
2007-07-20
Date posted
2007-09-08
Date terminated
2010-03-17
Location
Norcross, GA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.