Atlas FDA

BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)

FDA device recall Z-0602-2026 · posted 2025-11-26 · Open, Classified

Recall details

Recalling firm
BioFire Diagnostics, LLC
Reason for recall
Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.
Action taken
On May 28, 2026, Biomerieux issued a "Urgent: Medical Device Recall" Expansion Notification to affected consignees via E-Mail. Initial notification was sent out on October 22, 2025. Biomerieux asked consignees to take the following actions: 1. Examine your inventory for the lot identified in this field safety notice. 2. Discontinue use and discard any remaining product from this lot in your possession. 3. Please confirm the quantity of kits that have been destroyed, previously used, or returned on the attached Acknowledgement of Receipt Form. 4. If you have further distributed this product, please identify any recipients and notify them at once. 5. Please complete the accompanying Acknowledgement of Receipt Form and return to bioMrieux so that bioMrieux may acknowledge your receipt of this notification. 6. Complete and submit the report online. Regular Mail or Fax: Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
QOF
Quantity affected
97 Kits (2,910 pouches)
Distribution
US:AZ, FL,GA,ID NC, NH, NJ, NY, OR, PA, SC OUS: None
Date initiated
2025-10-22
Date posted
2025-11-26
Location
Salt Lake City, UT