Bio-logic-Ceegraph/Sleepscan Netlink Traveler, Bio-logic Systems Corp., a nebus company, Mundelein, IL 60060; Model 580-

FDA device recall Z-0761-2007 · posted 2007-04-20 · Terminated

Recall details

Recalling firm
Bio-Logic Systems Corp
Reason for recall
Battery and Battery Compartment; incorrect placement in the Netlink Traveler unit which caused a short in the wiring of the battery pack and overheating.
Action taken
On 2/26/07, Bio-logic Systems shipped replacement batteries, marked with part number 520-LIBAT4-3 and a label reading 'THIS SIDE OUT', and a new battery compartment cover with a label inside the cover to clarify the correct installation of the battery to each of the affected customers. The accounts were requested to remove and replace the old batteries, part 520-LIBTA4-2, and compartment covers, and return the old batteries and covers to Bio-logic Systems in the enclosed pre-paid shipping container, along with the completed accounting form.
Root cause
Other
Status
Terminated
Product code
GWQ
Quantity affected
130 battery packs (112 domestic USA, 18 foreign)
Distribution
Worldwide, USA , Canada, China, Hong Kong, India and Israel.
Date initiated
2007-02-26
Date posted
2007-04-20
Date terminated
2008-08-26
Location
Mundelein, IL