BinaxNOW Nasopharyngeal (NP) Accessory Pak, Part Number 400-065, contain NP Swabs with '8995' labeled on the swab wrappe

FDA device recall Z-0825-2007 · posted 2007-05-23 · Terminated

Recall details

Recalling firm
Binax, Inc. dba IMPD
Reason for recall
Tip of the foam swab could detach from the shaft causing injury to the patient.
Action taken
Inverness Medical issued recall notification by letter on 3/28/07. Customers are instructed to discard the NP swabs provided in the kits. The firm will replace the NP swabs. In the meantime customers are to use the tests with nasal wash/aspirate samples (influenza) or nasal wash samples (RSV). Response letters are to be signed by customers and returned to Inverness Medical.
Root cause
Other
Status
Terminated
Product code
GNX
Quantity affected
1101 kits
Distribution
Nationwide and Japan.
Date initiated
2007-03-28
Date posted
2007-05-23
Date terminated
2008-09-08
Location
Scarborough, ME