BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System.

FDA device recall Z-0193-2007 · posted 2006-11-16 · Terminated

Recall details

Recalling firm
Trionix Research Laboratory Inc
Reason for recall
The ball screw on the unit which supports the detector head (which weighs approximately 800 lbs.) could snap resulting in the detector head dropping on the patient.
Action taken
The recalling firm modified one of the devices in 2002. The remaining device users were telephoned on/about 11/02/2006. To date the recalling firm has not taken any corrective action regarding devices with serial numbers B-102, B-103, B-104 or B-106 thru B-184.
Root cause
Other
Status
Terminated
Product code
KPS
Quantity affected
76 devices
Distribution
Worldwide, including USA, Australia, Egypt, UK, Belgium.
Date initiated
2006-08-01
Date posted
2006-11-16
Date terminated
2008-10-29
Location
Twinsburg, OH