Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical aspiration of the uterine cavity.
FDA device recall Z-0078-2023 · posted 2022-10-12 · Open, Classified
Recall details
- Recalling firm
- Olympus Corporation of the Americas
- Reason for recall
- Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters which were not designed for use with the VC-10, which can lead to serious injuries as it does not fit properly with the VC-10.
- Action taken
- On September 16, 2022, the recalling firm notified affected customers through "URGENT Medical Device Field Corrective Action" letters. Customers were informed that the following new warning will be added to the IFU: Do not use replacement parts or accessories that are not compatible with the VC-10 system. Use of noncompatible accessories could lead to reduced system functionality and/or patient injury such as perforation, bleeding, and may require additional intervention. Refer to Section 11 for all accessories compatible with the VC-10 system. Customers should notify their users about risks associated with using non-compatible collection canisters and/or accessories and remind them to follow the list of compatible accessories outlined in Section 11 of the IFU. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. " Go to https://olympusamerica.com/recall. Enter the recall number 0414 " Enter your name and mailing address. Olympus will supply a copy of the revised IFU at a later date If you have distributed these devices outside your facility, please notify your customers of this matter immediately. If you have any questions, call (647) 999-3203 or email: Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
- Root cause
- Labeling design
- Status
- Open, Classified
- Product code
- HHK
- Quantity affected
- 4184 US; 191 OUS
- Distribution
- Domestic distribution nationwide. Foreign distribution to Canada and Japan.
- Date initiated
- 2022-09-16
- Date posted
- 2022-10-12
- Location
- Center Valley, PA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| Berkeley VC-10 Vacuum Curettage System (K171440) | Gyrus Acmi | 2017-09-29 |