Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800 Modification (MOD ID

FDA device recall Z-1184-2008 · posted 2008-08-20 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc
Reason for recall
Incorrect results: There is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range.
Action taken
Beckman Coulter mailed a Product Corrective Action (PCA) letter the week of October 8, 2007 to the customers who have the affected UniCel Dxl 800 Access lmmunoassay systems advising that there is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range. The letter further advises that Beckman Coulter will contact the customer to arrange a date to remove the modificatlons but until the modification is completed, the system should be operated in a room temperature of 20C to 30C. This can be acomplished by monitoring and recording room tamperature in the area of the instrument. Customers are also asked to review historical quality control records to ensure assay performance is within their laboratory specification and that patient results should be reviewed in context of other diagnostics tests and the clinical presentations of the patient. A response form is included.
Root cause
Device Design
Status
Terminated
Product code
JJE
Quantity affected
111 units in the US; 9 units in Canada
Distribution
Nationwide and Canada
Date initiated
2007-10-08
Date posted
2008-08-20
Date terminated
2012-01-11
Location
Brea, CA