BECKMAN COULTER MicroScan MICroSTREP Plus Panel, MicroScan Mueller-Hinton Broth with 3% lysed horse blood, Catalog Numbe

FDA device recall Z-1804-2018 · posted 2018-10-02 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
Beckman Coulter has received customer complaints of lowered minimum inhibitory concentration (MIC) for antimicrobial agents Cefaclor (Ctr) and Penicillin (P) with control organism Streptococcus pneumoniae (ATCC 49619) during quality control (QC) testing on MICroSTREP plus(R) Panel Type 1 hydrated with Mueller-Hinton Broth with 3% lysed horse blood part number B1015-25 lot number 1005986140.
Action taken
Beckman Coulter sent an Urgent Medical Device Recall letter dated March 16, 2018. Please discontinue use and discard any remaining inventory of the broth lot listed above. You should retain your inventory of other lot numbers of this broth as they are not impacted by this issue. Follow the guidance in the Broth Preparation section of the panel procedural and QC manual. Beckman Coulter recommends discussing the content of this letter with your medical director regarding the need to review previous test results, conduct patient follow-up, and/or repeat testing by another lot of this broth if the isolates are still available. For further questions, please call (469) 961-4941.
Root cause
Unknown/Undetermined by firm
Status
Terminated
Product code
JWY
Quantity affected
523 boxes
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Canada, Kuwait, South Korea and Saudi Arabia.
Date initiated
2018-03-26
Date posted
2018-10-02
Date terminated
2018-12-19
Location
Brea, CA