Beckman Coulter, Cholesterol Assay, Catalog No. OSR6X16
FDA device recall Z-1025-2015 · posted 2015-01-28 · Terminated
Recall details
- Recalling firm
- Beckman Coulter Inc.
- Reason for recall
- N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. NAC interference may lead to falsely low results.
- Action taken
- A customer notification letter was sent to their customers on 12/15/14 to inform them that Beckman Coulter has recently confirmed that N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations (for the treatment of acetaminophen overdose), has been determined to interfere with assays for the following analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to call (800) 223-0130 in the US and Canada. Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.
- Root cause
- Under Investigation by firm
- Status
- Terminated
- Product code
- CHH
- Quantity affected
- 59,965 units total (21,782 units in US)
- Distribution
- Nationwide in US and Puerto Rico. World Wide: Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China
- Date initiated
- 2014-12-15
- Date posted
- 2015-01-28
- Date terminated
- 2016-02-19
- Location
- Brea, CA