Atlas FDA

Beckman Coulter Bicarbonate, REF: OSR6237, 4x50 mL

FDA device recall Z-1607-2023 · posted 2023-05-18 · Open, Classified

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.
Action taken
On April 5, 2023, Beckman Coulter (BEC) issued a "URGENT MEDICAL DEVICE RECALL" Notification via mail and/or E-Mail. In addition, to informing consignees about the recalled product, BEC asked consignees to take the following actions: 1. Safely dispose of the Bicarbonate lot numbers that are specifically described in Table A. 2. If replacement or re-imbursement of the above lots is required, customers are directed to contact their local Beckman Coulter representative for replacement and re-imbursement. 3. Customers are advised to implement the adjusted Calibration OD range using the instructions within the customer letter to avoid generation of elevated results for all other Bicarbonate lots. 4. Retrospective review of patient data is not required due to the remote possibility of generating falsely elevated, clinically impactful results which could cause a change in treatment. Please share the content of this letter with your laboratory and/or medical director regarding the need to review previous patient test results. 4. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forward any of the affected products(s) listed above to another laboratory, please provide them a copy of the letter. 5. Please respond within 10 days in one of the following ways: - Electronically, you received this communication via email. - Manually, complete and return the enclosed Response Form. 6. If you have any questions, please call 1-800-854-3633
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
KHS
Quantity affected
48,633 units in total
Distribution
Worldwide Distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI,
Date initiated
2023-04-05
Date posted
2023-05-18
Location
Brea, CA