Atlas FDA

BD Visitec High Viscosity Injector 4mm [Hammer] .90 (20G) Catalog Number: 585208 High Viscosity Injectors are used durin

FDA device recall Z-2161-2011 · posted 2011-05-10 · Terminated

Recall details

Recalling firm
Becton Dickinson and Company
Reason for recall
Restricted or no fluid flow for Visitec High Viscosity Injector.
Action taken
Beaver Visitec (BVI) notified customers via telephone and by sending an Urgent Recall Customer Notification letter dated December 7, 2010. The letter identified the product, the problem, and the action to be taken by the customer. , Customers were requested to return all unused product in their inventory for credit or refund and complete the attached Customer Response Form and return it by fax to 781-893-7957. Questions can be directed to Inside Sales Manager at 781-906-7952.
Root cause
Component change control
Status
Terminated
Product code
HMX
Quantity affected
7614 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Israel, Jordan, Russia, Saudi Arabia, South Africa, Turkey, Iraq, Lebanon, Morocco, Tunisia, Syria, India, Argentina, Australia, Austria,
Date initiated
2010-12-17
Date posted
2011-05-10
Date terminated
2013-07-03
Location
Waltham, MA