BD Vacutainer SST Plus Blood Collection Tubes, 5.0 mL, 13 x 100 mm, Sterile. Provides a means of collecting, transportin

FDA device recall Z-2434-2015 · posted 2015-08-20 · Terminated

Recall details

Recalling firm
Becton Dickinson & Company
Reason for recall
Some of the tubes were manufactured with stoppers that did not meet current manufacturing specifications. Stoppers were manufactured with a reduced cycle cure time.
Action taken
Becton Dickinson notified their affected customers of this recall by sending them an Urgent Product Recall Letter dated June 23, 2015 via UPS 2nd Day.
Root cause
Process control
Status
Terminated
Product code
JKA
Quantity affected
238, 000 (Lot #5133959) and 243,000 (Lot #5133957)
Distribution
Distributed in FL, IL, MA, MI, MO, PA and TN.
Date initiated
2015-06-23
Date posted
2015-08-20
Date terminated
2016-04-06
Location
Franklin Lakes, NJ