BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352;
FDA device recall Z-1223-2024 · posted 2024-08-20 · Open, Classified
Recall details
- Recalling firm
- CareFusion 303, Inc.
- Reason for recall
- Automated dispensing cabinets, running affected software, and Windows 10 operating system may experience the application randomly hang, freeze and/or crash, which may result in delays in medication access, and potential data loss.
- Action taken
- On 2/1/24, correction notices were mailed or emailed to customers and distributors who were told the following: Devices with a random application hang, freeze and/or crashing can be resolved with a reboot of the device. In downtime scenarios, medications can be obtained from other devices or from pharmacy. There are also keys that can be used to manually access contents within the device. 1) Ensure you continue to manually reboot your devices as a part of your maintenance schedule. 2. Disseminate this notification within your facility network to ensure they are aware of the issue. 3. If you experience this issue, please contact Technical Support. 4. Complete the Customer Response Form and return to BDRC53@bd.com If you require further assistance, please contact the firm: Customer Support: 1-866-583-8783, 5:00am PST to 4:00pm PST (8:00am ET to 7:00pm ET), Monday - Friday. Technical Support: bd.com/self-service, 1-800-727-6102, 24 hours, 7 days a week
- Root cause
- Software design
- Status
- Open, Classified
- Product code
- BRY
- Quantity affected
- 20,443
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of VA, SC, IL, AL, KS, LA, GA, FL, CA, PR, PA, WV, WA, HI, NC, MD, AR, NY, ND, TX, NM, GU, MS, AZ, DC, AK, NE, DE, CO, SD, M
- Date initiated
- 2024-02-01
- Date posted
- 2024-08-20
- Location
- San Diego, CA