BD Pyxis MedStation ES AUX, REF 324
FDA device recall Z-0972-2025 · posted 2025-01-17 · Open, Classified
Recall details
- Recalling firm
- CareFusion 303, Inc.
- Reason for recall
- Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,
- Action taken
- BD issued an Urgent Medical Device Product Advisory to its consignees on 02/14/2023 via letter. The notice explained the problems with the device, health hazards, and requested the following actions be taken: "1. Customers may continue to use these devices as intended and should file a complaint via the Customer Portal at https://www.bd.com/self-service or call BD Support at 1-800-727-6102, should you experience drawer or door failures. 2. If you experience delays in access to medication due to a drawer or door failure, ensure backup procedures (which may include but may not necessarily be limited to keys, emergency medication supplies, alternate devices in proximity, etc.) are in place and staff have all been properly trained on them. 3. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter. 4. If you have specific questions about this letter, please contact 1-866-583-8783." The firm issued a second notice on 01/09/2024 with the same information.
- Root cause
- Nonconforming Material/Component
- Status
- Open, Classified
- Product code
- BRY
- Quantity affected
- 39615 units
- Distribution
- Worldwide
- Date initiated
- 2023-02-14
- Date posted
- 2025-01-17
- Location
- San Diego, CA