BD Pyxis Anesthesia ES System (PAS ES), REF: 327; BP Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis MedStation ES
FDA device recall Z-1207-2023 · posted 2023-03-17 · Open, Classified
Recall details
- Recalling firm
- CareFusion 303, Inc.
- Reason for recall
- Automated dispensing cabinet drawer firmware may have inadvertently been downgraded to an earlier version of drawer firmware during new device installations or device replacement upgrades, which may lead to an increased frequency of drawer failures, which could increase the likelihood of patients experiencing delays in receiving medications.
- Action taken
- On 01/14/23, correction notices were mailed to directors of pharmacy informing them to take the following actions: 1) Ensure this letter and contents are forwarded and/or discussed with all appropriate personnel to ensure proper awareness of the issue. 2) If you are experiencing any issues, such as drawer failures, call BD technical support and refer to user guides for additional information and troubleshooting procedures. a. BD Pyxis MedStation ES user guide, refer to sections "Recover storage space" (Chapter 9, pg. 76) and "Emergency medication removal" (Appendix 11, pg. 119). See references 1 and 2 below for additional details. b. BD Pyxis PAS ES user guide, refer to sections "Emergency access procedures" (Appendix A, pg. 77) and "Storage spaces are not responsive" (Appendix B, pg. 81). See references 3 and 4 below for additional details. 3) Complete and return the Customer Response Form. The firm will take the following actions: 1) All customers who may have affected devices will be notified. 2) Upon receipt of completed response forms, the firm will coordinate the firmware upgrade with customers. Technical Support: 1-800-727-6102, Phone hours: 24 hours, 7 days a week
- Root cause
- Software change control
- Status
- Open, Classified
- Product code
- BRY
- Quantity affected
- 7914
- Distribution
- Worldwide - US Nationwide distribution including in the states of NY, PA, OH, TX, FL, KY, IL, DE, AK, WI, KS, LA, WA, AZ, CO, MI, IN, TN, SD, AR, MO, CA, NV, MA, SC, VA, WV, CT, ME, NE, RI, GA, NC, IA
- Date initiated
- 2023-02-14
- Date posted
- 2023-03-17
- Location
- San Diego, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.