BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx Only. The f

FDA device recall Z-2598-2018 · posted 2018-07-13 · Terminated

Recall details

Recalling firm
Becton, Dickinson and Company, BD Biosciences
Reason for recall
The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.
Action taken
The firm, BD Life Sciences, produced the "URGENT PRODUCT RECALL-CORRECTION" letter dated 4/6/2018 to its customers during an on-site visit. by the field service engineer made beginning 5/16/2018 to correct the issue. The visits were completed 6/15/2018. If you require further assistance, contact BD Customer/Technical Support BD Customer Support at 877.232.8995 (prompt 2) for instruments in the United States. For customers outside the US, contact your local BD representative or distributor.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
OYE
Quantity affected
8 devices
Distribution
US Distribution was made to FL, KS, KY, MA, NJ, and SD.
Date initiated
2018-05-16
Date posted
2018-07-13
Date terminated
2018-08-13
Location
San Jose, CA