Atlas FDA

BD FACS Sample Prep Assistant II, Part Number: 337170; Product is manufactured and distributed by BD Biosciences, San Jo

FDA device recall Z-0526-2008 · posted 2008-04-10 · Terminated

Recall details

Recalling firm
BD Biosciences
Reason for recall
Mixing Performance: SPA II devices manufactured with the affected product motor control board may exhibit suboptimal mixing performance.
Action taken
On 10/1/07, all consignees were notified via Important Product Information letter and also an on-site visit by the firm's sales representative. The letter informed users of the affected product and provided instructions on the recall. For additional concerns or questions contact BD Customer Support Center at 1-877-232-8995.
Root cause
Software change control
Status
Terminated
Product code
GKI
Distribution
Worldwide Distribution --- USA including states of: NJ, IL, TX, PA, NY, MN, and AZ, and countries of Australia, Belgium, Hong Kong, Trinidad and Tobago. Australia, Belgium, Hong Kong, Trinidad and Tob
Date initiated
2007-10-02
Date posted
2008-04-10
Date terminated
2010-12-13
Location
San Jose, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
MODIFICATION TO BD FACSCANTO SYSTEM WITH BD FACSCANTO CLINICAL SOFTWARE (K050191)Becton, Dickinson & CO2005-02-22