BD Biosciences MultiSET Flow Cytometry Software Versions 2.1, 2.1.1, and 2.2 with OS X operating systems, BD Biosciences

FDA device recall Z-0570-2008 · posted 2008-03-04 · Terminated

Recall details

Recalling firm
BD Biosciences
Reason for recall
Inaccurate results: Software error results in inaccurate display result statistics. If the user adjusts the lymph gate or attractors in the Lab Report view without using the Manual Gate function (as described in the system manual), the statistic results will not be updated to match the adjusted gate.
Action taken
Becton Dickinson sent an Important Product Information letter by mail on 09/14/2007 informing customers of the need to follow the instructions for adjusting the lymph gate or attractors. The next software revision will alleviate this limitation.
Root cause
Software design
Status
Terminated
Product code
GKZ
Quantity affected
1136 units
Distribution
Worldwide Distribution - USA, Belgium, Brazil, Peru, Singapore, Canada, Jamaica, Chile, Ecuador, Hong Kong-China, Thailand, Mexico, Venezuela, Barbados, Philippines, Puerto Rico, Guatemala, Japan, Uru
Date initiated
2007-09-06
Date posted
2008-03-04
Date terminated
2008-04-15
Location
San Jose, CA