BD BACTEC FX 40 instrument, catalog number 442296, manufactured by BD Diagnostic Systems, 7 Loveton Circle, Sparks, MD 2
FDA device recall Z-0870-2014 · posted 2014-01-30 · Terminated
Recall details
- Recalling firm
- Becton Dickinson & Co.
- Reason for recall
- Improperly functioning component of a diagnostic medical device may cause false negative results in clinical specimens.
- Action taken
- BD Diagnostics initiated a field correction by phone on December 9, 2013, to affected customers and followed the calls with letters, e-mails, or fax. BD Diagnostics sent a secondary Urgent Field Corrective Action letter dated November 2013 to all affected customers. The recalling firm Secondary notifications were serial number specific and explained that some BD BACTEC FX instruments contain racks that may demonstrate variable indicator light intensities. Specifically, the Anonymous vial indicator lights which should appear yellow may instead appear more green in color. BD noted that affected instruments continue to function properly and will not report an incorrect result. The issue is limited to the yellow indicator lights only. Red and green indicator lights appear as intended. BD is developing software to correct this issue and anticipates that the new software will be available to begin instrument updates by February, 2014. The new software will not affect instrument if the indicator lights are functioning as intended. Until the software can be updated on your instrument, BD provided workflow instructions to ensure that the indicator light issue does not occur. For questions regarding this recall call 410-316-4054.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- MDB
- Quantity affected
- 2 units
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to IN, AU, CN, AR, BE, PE, KR, BR, CO, SG, TH, BN, HK, JP, and CA.
- Date initiated
- 2013-12-09
- Date posted
- 2014-01-30
- Date terminated
- 2014-08-20
- Location
- Sparks, MD