Atlas FDA

BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions

FDA device recall Z-2179-2025 · posted 2025-08-06 · Open, Classified

Recall details

Recalling firm
CareFusion 303, Inc.
Reason for recall
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Action taken
On December 12, 2025, CareFusion 303, Inc. (BD) issued a updated "Urgent Medical device Recall Notice via E-Mail and physical mail. The updated notice includes important information on infusion sets as well as labeling updates. Previous notifications were sent on July 8,2025, July 17, 2025, September 11, 2025 and September 12, 2025. CareFusion ask consignees to take the following actions: 1. Review your inventory for the catalog numbers included in the updated letter. Remove and discard any remaining inventory per your facility guidelines. 2.To receive credit or replacement product, please return the completed customer response form attached to this notice. 3.Please contact your BD Clinical Consultants and Account Executives for alternative sets information. 4. Review the updated sets performance data included in the letter. 5. Circulate this notice within your facility network to ensure that all concerned personnel are made aware of this issue. 6. Distributors should provide a copy of this notice to all customers who may have purchased an Alaris Pump Module. If you require further assistance, please contact: BD Technical Support at Phone: 888-812-3229; Phone hours: 6:00am to 5:00pm PT Monday Friday and Email: DL-US-INF-TechSupport@bd.com.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
FRN
Quantity affected
33,162,680 units total
Distribution
US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MP MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI
Date initiated
2025-07-08
Date posted
2025-08-06
Location
San Diego, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
BD Alaris System with Guardrails Suite MX v12.1.2 (K211218)Carefusion 303, Inc.2023-07-21