Atlas FDA

BD Alaris PCU REF 8015

FDA device recall Z-0184-2024 · posted 2023-11-07 · Open, Classified

Recall details

Recalling firm
CareFusion 303, Inc.
Reason for recall
Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Action taken
On September 15, 2023, CareFusion a subsidiary of Becton Dickinson (BD) posted an "Urgent Medical Device Labeling Correction" on their website. On October 4, 2023 CareFusion mailed a copy of the correction notification via FedEx. CareFusion asked consignees to take the following actions: 1. You should cease use of Cardinal Health branded Monoject" syringes with the BD Alaris" Syringe and PCA Modules; except for SKU 8881135609, which is not impacted by this notification. 2. Please refer to Attachments 1 and 2 of this notice for a full selection of syringes that are compatible with the BD Alaris" Syringe and PCA Modules. 3. Replace your current BD Alaris" Syringe and PCA Module Compatibility Lists with Attachments 1 and 2 of this notice. 4. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred BD Alaris" Infusion System devices to. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
FRN
Quantity affected
867,362 units
Distribution
Worldwide Distribution: US (nationwide): AK, AL, AR, AS, AZ, CA, CO, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR,
Date initiated
2023-09-15
Date posted
2023-11-07
Location
San Diego, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
BD Alaris System with Guardrails Suite MX v12.1.2 (K211218)Carefusion 303, Inc.2023-07-21
ALARIS SYSTEM WITH GUARDRAILS SUITE MX (K133532)Carefusion 303, Inc.2014-08-21
ALARIS SYSTEM WITH MMS, MODEL 8XXX (K072105)Cardinal Health 303,Inc.2007-12-28
ALARIS SYSTEM PC UNIT, MODEL 8001 (K051641)Cardinal Health, Alaris Products2005-10-20