Bayer Contour TS blood glucose test strips packaged in packs of 10, 25, 50 and 100 strips; Bayer Healthcare, Mishawaka,

FDA device recall Z-0932-2008 · posted 2008-03-13 · Terminated

Recall details

Recalling firm
Bayer Healthcare LLC
Reason for recall
Inaccurate Test Strips Results: Results in blood glucose readings with a positive bias are outside the product specifications. Patient test results may demonstrative results 5 percent to 17 percent higher.
Action taken
The firm issued a Press Release on 12/21/07. A Test Strip Recall letter was also sent to the firm's consignees and health care providers instructing them that use of the product should cease and that health care providers contact the firm for replacements. Additional information can be found at www.bayerdiabeties.com or by contacting the Bayer at 1-800-348-8100.
Root cause
Process design
Status
Terminated
Product code
LFR
Quantity affected
97,000 bottles of test strips
Distribution
Worldwide Distribution --- including USA and countries of Austria, France, India, Korea, Mexico, Philippines and Turkey.
Date initiated
2007-12-21
Date posted
2008-03-13
Date terminated
2009-07-07
Location
Mishawaka, IN