Bayer Advia 70 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX.

FDA device recall Z-0448-2007 · posted 2007-01-09 · Terminated

Recall details

Recalling firm
Drew Scientific Group PLC
Reason for recall
Flags and alerts on the user interface will not automatically transfering to the LIS.
Action taken
Firm sent 'Product Notification #22' to all customers on 08/24/06. Users advised of potential risk to patient health and provided with instructions on how to eliminate risk by disabling the auto transfer feature and manually transfering data until a software patch is made available
Root cause
Other
Status
Terminated
Product code
GKZ
Quantity affected
545 units.
Distribution
Worldwide distribution --- including USA states of FL, MA, MN, ND, NM, NY, Puerto Rico and TX and countries of Austria, Bangladesh, Chile, China, Czech Republic, Egypt, France, Hong Kong, Italy, India
Date initiated
2006-08-24
Date posted
2007-01-09
Date terminated
2008-03-10
Location
Dallas, TX