Atlas FDA

Baxter Interlink System Continu-Flo Solution Set with Standard Bore 4-Way Stopcock Manifold Extension Set and Minivolume

FDA device recall Z-0653-06 · posted 2006-03-24 · Terminated

Recall details

Recalling firm
Baxter Healthcare Corp.
Reason for recall
The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
Action taken
Baxter sent Urgent Product Recall letters dated 8/3/05 to their customers, to the attention of the Director of Materials Management, via first class mail on the same date. The accounts were informed of the microbore female luers that may crack during use, resulting in the possible under delivery or incomplete delivery of medication. The accounts were instructed to examine their inventory for the affected lots of product, remove it from inventory and return it to Baxter for credit. Any questions were directed to the Center for One Baxter at 1-800-422-9837. Baxter extended the recall to include additional lots and product codes via letter dated 12/15/05. The letters referenced the 8/3/05 letter, including a copy of that letter, and listed the affected lots of sets. The instructions for return of the product remained the same as that described in the 8/3/05 letter.
Root cause
Other
Status
Terminated
Product code
FMG
Quantity affected
1,920 sets
Distribution
Nationwide, including Puerto Rico and American Samoa, and internationally to Canada, Australia, Japan, South Korea, Saudi Arabia, India, Mexico, Venezuela, Malaysia, East Timor and New Zealand.
Date initiated
2005-12-15
Date posted
2006-03-24
Date terminated
2008-08-18
Location
Round Lake, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ELCAM STOPCOCKS AND MANIFOLDS (K022895)Medical Device Consultants, Inc.2002-10-18
STOPCOCK MANIFOLD GANGS (K962581)Baxter Healthcare Corp1996-08-28
CONTINUFLO SOLUTION SET/SECONDARY MEDICATION SET (K961225)Baxter Healthcare Corp1996-06-21
INTERLINK SYSTEM - T-CONNECTOR EXTENSION SET (K921899)Baxter Healthcare Corp1993-04-20