Baxter Interlink System Continu-Flo Solution Set, length 89" (2.3 m), 2 Injection Sites, Male Luer Lock Adapter, 3-Port
FDA device recall Z-0478-2012 · posted 2012-01-11 · Terminated
Recall details
- Recalling firm
- Baxter Healthcare Corp.
- Reason for recall
- During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection.
- Action taken
- The firm, Baxter, sent an "Urgent Product Recall" letter dated December 6, 2011, via first class mail to its customers. The letter described the product, problem, and actions to be taken. The customers were instructed to inspect all locations where their facility's inventory is located for products with the affected lot numbers; return the affected product; request a credit or replacement by calling Baxter Healthcare Center for Service at 1-888-229-0001 between 7:00am to 6:00pm Central time, Monday through Friday; complete and return the attached Customer Reply Form via fax to: 1-847-270-5457, or scan and email to fca@baxter.com; and if they are a dealer, wholesaler, or distributor/reseller that distributed any affected product to other facilities, please notify their customers of this recall. If you have any questions regarding the communication, please call The Center for One Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CST.
- Root cause
- Employee error
- Status
- Terminated
- Product code
- FMG
- Quantity affected
- 1,488 units
- Distribution
- Worldwide distribution: USA (nationwide) including states of: Alaska, California, Colorado, Florida, Georgia, Hawaii, Illinois, Kansas, Kentucky, Massachusetts, Maine, Michigan, Minnesota, Missouri, N
- Date initiated
- 2011-12-06
- Date posted
- 2012-01-11
- Date terminated
- 2012-11-01
- Location
- Round Lake, IL
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| MULTI-PORT MANIFOLD (K932512) | Baxter Healthcare Corp | 1994-02-22 |